引用本文
  • 李桂香,胡湛波,柴欣生.悬浮物对生活污水中总氮与氨氮浓度检测的影响[J].广西科学院学报,2022,38(3):311-318.    [点击复制]
  • LI Guixiang,HU Zhanbo,CHAI Xinsheng.Study on the Effect of Suspended Solids on the Detection of Total Nitrogen and Ammonia Nitrogen Concentration in Domestic Sewage[J].Journal of Guangxi Academy of Sciences,2022,38(3):311-318.   [点击复制]
【打印本页】 【在线阅读全文】【下载PDF全文】 查看/发表评论下载PDF阅读器关闭

←前一篇|后一篇→

过刊浏览    高级检索

本文已被:浏览 194次   下载 191 本文二维码信息
码上扫一扫!
悬浮物对生活污水中总氮与氨氮浓度检测的影响
李桂香1, 胡湛波1, 柴欣生2
0
(1.广西大学资源环境与材料学院, 广西南宁 530004;2.华南理工大学制浆造纸工程国家重点实验室, 广东广州 510641)
摘要:
针对标准方法检测实际生活污水水样时经常会出现氨氮(NH3-N)浓度大于总氮(TN)浓度的不合理现象,本研究探讨了生活污水水样中悬浮物(SS)浓度及过滤处理方式对检测结果的影响。结果表明,SS浓度与标准方法测定偏差呈正比,原水样和0.45 μm膜过滤处理的水样TN浓度测定结果与实际偏差高达11.1%,滤纸与0.45 μm膜过滤处理的水样NH3-N浓度测定结果与实际偏差高达31.4%。此外,标准方法和改进方法的测定偏差与SS浓度的线性关系良好,经验校正方程的提出减少了SS带来的测定误差。因此,采用标准方法检测生活污水水样TN和NH3-N浓度前建议使用0.45 μm滤膜过滤预处理,或采用本研究提出的经验校正方程进行校正,以提高标准方法检测的准确性。
关键词:  悬浮物  总氮  氨氮  标准方法  改进方法
DOI:10.13657/j.cnki.gxkxyxb.20221019.013
投稿时间:2021-12-16修订日期:2022-02-26
基金项目:广西重点研发计划(桂科AB19110001)资助。
Study on the Effect of Suspended Solids on the Detection of Total Nitrogen and Ammonia Nitrogen Concentration in Domestic Sewage
LI Guixiang1, HU Zhanbo1, CHAI Xinsheng2
(1.School of Resources, Environment and Materials, Guangxi University, Nanning, Guangxi, 530004, China;2.State Key Laboratory of Pulp and Paper Engineering, South China University of Technology, Guangzhou, Guangdong, 510641, China)
Abstract:
In view of the unreasonable phenomenon that the concentration of ammonia nitrogen (NH3-N) is often greater than the concentration of total nitrogen (TN) when the standard method is used to detect the actual domestic sewage water samples,the effects of suspended solids (SS) concentration and filtration treatment on the detection results of domestic sewage samples were studied in this study.The results showed that the concentration of SS was proportional to the determination deviation of the standard method.The deviation of TN concentration between raw water samples and water samples treated by 0.45 μm membrane filtration was as high as 11.1%.The deviation of NH3-N concentration in water samples treated with filter paper and 0.45 μm membrane filtration was as high as 31.4%.In addition,the measurement deviation of the standard method and the improved method has a good linear relationship with the SS concentration,and the proposal of the empirical correction equation reduces the measurement error caused by SS.Therefore,before using the standard method to detect the TN and NH3-N concentrations in domestic sewage water samples,it is recommended to use 0.45 μm filter membrane filtration pretreatment,or use the empirical correction equation proposed in this study to improve the detection accuracy of the standard method.
Key words:  suspended solids  total nitrogen  ammonia nitrogen  standard method  improved method

用微信扫一扫

用微信扫一扫